FaecoCheck test is able to distinguish the bleeding from Lower GIT or Upper GIT in the routine screening.
Cat. No: VH-010 (100 tests/kit) - Patent Pending
Lateral Flow, Membrane Based Monoclonal Immunoassay for Qualitative Detection of Human Haemoglobin in Faeces
INTENDED USE
The Hanson® Faecocheck Test is an easy to use, contains of both immuno-test and chemical test. The monoclonal immunoassay is major for the Lower GIT screening and designed to provide a qualitative analysis of human haemoglobin in faecal samples. The chemical test is major for the Upper GIT screening and designed to provide a qualitative analysis of human heme in faecal samples. The Faecocheck Test combined immunoassay and chemical test in the same device as an all-in-one faecal occult blood screening test. The test is intended for use by medical laboratories and physicians. The assay should not be attempted without proper supervision and is not intended for over the counter sale to laypersons.
The Hanson® Faecocheck Test is a screening test providing a preliminary analytical test result only. It should be used in conjunction with other tests to properly exclude the presence of gastrointestinal tract (GIT) pathologens. Any positive results should be promptly followed up using more definitive tests such as colonoscopy or barium enema X-ray.
PRINCIPLE OF TEST
The Hanson® Faecocheck Test is an affordable, qualitative screening test used to identify individuals, asymptomatic or otherwise, for further investigative testing. The detection of faecal occult blood who will need a indicates GIT bleeding. Although some daily blood loss to the GIT is normal (1), excessive bleeding is an indicator of a range of pathological conditions from insignificant to potentially fatal. It is through the early detection of faecal occult blood that the effects and progress of serious pathologies such as polyps, diverticulae, ulcers and colorectal cancer may be reduced or removed completely.
The Hanson® Faecocheck Test combined immunoassay and chemical test in the same device. The immunological aspect of the Hanson® Faecocheck Test uses a membrane bound monoclonal anti-human haemoglobin antibody to specifically bind human haemoglobin in faecal samples. The specificity of the antibody for human haemoglobin eliminates the possibility of false positive results stemming from heme(hematin), dietary sources of haemoglobin and dietary peroxidases.
Once bound to the monoclonal antibody the hemoglobin’s pseudoperoxidase activity is utilized to produce a blue coloured end product. The reaction involves the oxidation of alpha guaiaconic acid by the catalytic action of haemoglobin in the presence of peroxide. The product of the reaction appears as a localized blue line on the test strip. The appearance of such product constitutes a positive reaction. If no haemoglobin is present in the sample or the amount present is less than the minimum detection limit of the test, no coloured end product will form and the result is deemed negative. The immunoassay is major for the Lower GIT screening although some of haemoglobin will through stomach into intestinal tract (2).
The chemical aspect of the Hanson® Faecocheck Test is designed to detect human heme that digests from upper GIT bleeding. It also uses the pseudoperoxidase activity of heme groups to produce a blue coloured end product. However, without the ability to selectively bind human haemoglobin, all haemoglobins (human or animal), partially denatured haemoglobins, and individual heme molecules will produce this coloured end product. This lack of specificity, which is so often the cause of false positives particularly when one uses the chemical testing method alone, is used by the Hanson® Faecocheck Test to indicate bleeding originating from the upper GIT. That is, by being able to detect the presence of heme groups and other products of human haemoglobin that have been partially digested by the stomach and the test can compare the bleeding originated from the Upper GIT with whole haemoglobin associated with lower GIT bleeds.
REAGENTS/MATERIALS PROVEDED
Each Hanson® Faecocheck Test kit contains:
WARNINGS AND PRECAUTIONS
FOR IN-VITRO DIAGNOSTIC USE ONLY
STORAGE
Hanson® Faecocheck Test kits are best stored at 2-8℃. (Storage at ambient room temperature less than 30℃ is also permissible for up to 2 weeks).
SPECIMEN COLLECTION AND HANDLING
If using the Applicator Stick to apply a faecal sample directly to the sample application well, best results are obtained using a faecal collector(container) or bag to collect the stool. Only a small sample of the stool is required for the test but it must be a representative sample. For best results the sample should be obtained from at least two parts of the stool, ideally the inner and outer areas to maximize the likelihood of detecting any blood.
If using the dropper to apply an aqueous sample, simply draw up a dropper full of water (distilled water) into the faecal container and mix well with stool specimen. Open flap A on the Faecocheck test slide, then adds 5 drops of the stool-water suspension solution into the sample application well. Allow time for all solution to be absorbed.
All samples should be applied to the Hanson® Faecocheck strip as soon as possible to avoid degradation of haemoglobin.
As GIT bleeding is often intermittent, three consecutive stool samples should be examined using one Hanson® Faecocheck slide per sample.
ASSAY PROCEDURES
A) AT THE HOME
PROCEDURE FOR USING THE APPLICATOR STICK:
Patient to complete steps 1 - 7; Physician to complete steps 8 - 12.
B) IN THE LABS
PROCEDURE FOR USING THE DROPPER:
Patient to complete steps 1 ; Physician to complete steps 2 - 9.
INTERPRETATION OF RESULTS
Results are to be read 3 to 5 minutes after the addition of hydrogen peroxide solution to the test strip. Any significance colour development occurring after 5 minutes is irrelevant and has no influence on the test result interpretation.
Immunological method (for Lower GIT Screening)
A positive result indicates the presence of human haemoglobin in the faecal sample and presents as the formation of a blue line in the test strip area.
A negative result is obtained when there is no haemoglobin in the sample or the amount present is below the lower detection limit of the test. No blue coloured end product forms in the test strip area when a result is negative.
Chemical method (for Upper GIT Screening)
A positive result indicates the presence of whole haemoglobin, components of haemoglobin an/or haem groups all of which may be of human or animal/dietary origin. A positive result is represented by a generalized distribution of blue coloured end product over the sample pad area as well as on the absorbent membrane in the sample application well.
A negative result is obtained when no haemoglobin, haemoglobin digests or haem groups are present in the sample or they are present at concentrations below the detection limit of the test. No blue coloured end product forms in the sample pad area for the negative sample.
1. | Immuno-(Negative) Test Strip has no blue line. Chemical-(Negative) Sample Pad is not forming blue colour. |
Fig. 1 | |
2. | Immuno-(Positive) Test Strip has a blue line. Chemical-(Positive) Sample Pad is forming blue colour. |
Fig. 2 | |
3. | Immuno-(Negative) Test Strip has no blue line. Chemical-(Positive) Sample Pad is forming blue colour. |
Fig. 3 | |
4. | Immuno-(Positive) Test Strip has a blue line. Chemical-(Negative) Sample Pad is not forming blue colour. |
Fig. 4 |
QUALITY CONTROL
Quality control testing should be carried out whenever a batch of patient samples is run. Hanson Hong Biomedical is able to supply the positive and negative controls. Alternatively one can use an in house positive control as follows:
LIMITATIONS
PERFORMANCE
The monoclonal antibody used in the Hanson® Faecocheck Test has been tested in-vitro against rabbit, pig, sheep, cow and fish haemoglobin with all producing negative results. Peroxidases derived from vegetables were tested and results were negative. This is due to the high specificity of the antibody for human haemoglobin.
The FaecoCheck test was compared with a commercial available immunoassay (OC-Hemodia, Eiken) for human haemoglobin which has a cut-off value at 0.2 - 0.4 m g/ml. 44 clinical stool samples were evaluated by both systems. Of the 44 samples, 5 were determined positive and 39 were determined negative by the OC-Hemodia method; 6 were determined as positive and 38 were determined as negative by FaecoCheck (immunoassay) method; 34 were determined as positive and 10 were determined as negative by FaecoCheck (chemical assay) method; all of 44 were determined as positive by O-toluidine method. Six positives were proved that is including colorectal cancer 1 case、malignant gastrointestinal tumor 1 case、duodenal ulcer 2 cases、acute gastritis 1 case and acute salmonella infected 1 case. The FaecoCheck test correctly identified 100 % of these positive samples and 100 % of these negative samples. It may be liquid faces' samples to occur 1 false negative from OC-Hemodia. All of 6 positive cases were determined positive by FaecoCheck (chemical assay) method.
REFERENCES
Cat. No. | Description | Packing |
VH-010 | FaecoCheck Test | 100 tests/kit |
VH-010-1 | Reagent 1 & 2 | pair |
VH-010-2 | Applicator Stick | 100ea/pk |
VH-010-3 | Dropper | 100ea/pk |
FC-010 | Positive control (standard Human Hb) | 2 ml/vial |
FC-011 | Faeces container | 1000ea/pk |